RT info:eu-repo/semantics/article T1 Signal transfer with excitation-emission matrices between a portable fluorimeter based on light-emitting diodes and a master fluorimeter A1 Sanllorente Méndez, Silvia A1 Rubio Martínez, Laura A1 Ortiz Fernández, Mª Cruz A1 Sarabia Peinador, Luis Antonio K1 Signal transfer K1 Portable fluorimeter K1 Excitation-emission fluorescence K1 PARAFAC K1 Capability of detection K1 Enrofloxacin K1 Química analítica K1 Chemistry, Analytic AB In this work, the transfer of the excitation-emission matrices between a portable fluorimeter based on LEDs and a master fluorimeter based on a xenon source was carried out. Enrofloxacin was the analyte of interest and it was measured alone or in binary mixtures with flumequine (partially overlapped signals) or with ciprofloxacin (fully overlapped signals). The maintenance and transfer of the unequivocal identification of the fluorophores between both instruments are shown. The precision in the determination performed with the portable fluorimeter approximated to that made with the master fluorimeter using this transfer and it did not introduce bias. The correlation coefficients of the calibrations based on PARAFAC using EEM signals were higher than 0.999, whereas the values of the capability of detection ranged from 14.8 to 26.9 μg L−1 for probabilities of false positive and false negative fixed at 0.05. These results contribute to the effort to perform the fluorimetric detection outside the laboratory and to promote the use of databases of fluorescence spectra for the unequivocal identification in remote of fluorophores of interest and/or regulated. PB Elsevier SN 0925-4005 YR 2019 FD 2019-04 LK http://hdl.handle.net/10259/5058 UL http://hdl.handle.net/10259/5058 LA eng NO Spanish MINECO (AEI/FEDER, UE) through projects CTQ2014-53157-R and CTQ2017‐88894‐R and by Junta de Castilla y León through project BU012P17 (all co‐financed with European FEDER funds) DS Repositorio Institucional de la Universidad de Burgos RD 25-abr-2024