dc.description.abstractIn this work, the transfer of the excitation-emission matrices between a portable fluorimeter based on LEDs and a master fluorimeter based on a xenon source was carried out. Enrofloxacin was the analyte of interest and it was measured alone or in binary mixtures with flumequine (partially overlapped signals) or with ciprofloxacin (fully overlapped signals). The maintenance and transfer of the unequivocal identification of the fluorophores between both instruments are shown. The precision in the determination performed with the portable fluorimeter approximated to that made with the master fluorimeter using this transfer and it did not introduce bias. The correlation coefficients of the calibrations based on PARAFAC using EEM signals were higher than 0.999, whereas the values of the capability of detection ranged from 14.8 to 26.9 μg L−1 for probabilities of false positive and false negative fixed at 0.05. These results contribute to the effort to perform the fluorimetric detection outside the laboratory and to promote the use of databases of fluorescence spectra for the unequivocal identification in remote of fluorophores of interest and/or regulated.en
dc.description.sponsorshipSpanish MINECO (AEI/FEDER, UE) through projects CTQ2014-53157-R and CTQ2017‐88894‐R and by Junta de Castilla y León through project BU012P17 (all co‐financed with European FEDER funds)en
dc.format.mimetypeapplication/pdf
dc.language.isoenges
dc.publisherElsevieren
dc.relation.ispartofSensors and Actuators B: Chemical. 2019, V. 285, p. 240-247en
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 International
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subjectSignal transferen
dc.subjectPortable fluorimeteren
dc.subjectExcitation-emission fluorescenceen
dc.subjectPARAFACen
dc.subjectCapability of detectionen
dc.subjectEnrofloxacinen
dc.subject.otherQuímica analíticaes
dc.subject.otherChemistry, Analyticen
dc.titleSignal transfer with excitation-emission matrices between a portable fluorimeter based on light-emitting diodes and a master fluorimeteren
dc.typeinfo:eu-repo/semantics/article
dc.date.embargo2021-04-15
dc.rights.accessRightsinfo:eu-repo/semantics/openAccess
dc.relation.publisherversionhttps://doi.org/10.1016/j.snb.2019.01.051
dc.identifier.doi10.1016/j.snb.2019.01.051
dc.relation.projectIDinfo:eu-repo/grantAgreement/MINECO/CTQ2014-53157-R
dc.relation.projectIDinfo:eu-repo/grantAgreement/MINECO/CTQ2017‐88894‐R
dc.relation.projectIDinfo:eu-repo/grantAgreement/JCyL/BU012P17
dc.type.hasVersioninfo:eu-repo/semantics/acceptedVersionen


Ficheros en este ítem

Thumbnail

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem